A Medical Device Daily
Medtronic (Minneapolis) reported Friday the start of an international clinical study to confirm the safety and efficacy of the EnRhythm MRI SureScan pacing system, the first-ever pacemaker system to be developed and tested specifically for safe use in magnetic resonance imaging (MRI) machines under specified scanning conditions.
The pacing system consists of the dual-chamber SureScan pacemaker and CapSureFix MRI SureScan pacing leads. The first clinical trial devices were recently implanted in Europe. All patients in the trial will receive an EnRhythm MRI SureScan pacing system, and approximately half of the participants will receive MRI scans; the trial's first MRI scans will occur approximately two to three months after device implant. The study will compare the electrical and diagnostic system performance, as well as complication rates, among patients in both groups. The expected study duration is approximately 24 months.
The EnRhythm MRI SureScan pacemaker includes modified hardware to minimize the level of energy transmitted through the lead/device connection point. The pacemaker also includes a new SureScan feature that can be programmed "on" before an MRI scan to eliminate the impact of MRI-generated electrical noise, which can prevent necessary pacing therapy or cause the device to oversense and deliver unnecessary pacing therapy. When the SureScan feature is on, the device's data collection and monitoring functions are temporarily suspended, while allowing the device to continue providing asynchronous pacing if needed.